The 56 is a line powered analyzer offering single or dual sensor input with an unrestricted choice of dual measurement combinations. The 56 measures pH, ORP, ISE, Conductivity, Chlorine, Dissolved and Gaseous Oxygen, Dissolved Ozone, Temperature, Turbidity, or Flow and will accept 4-20mA current input. The device is offered in 24VDC or 115/230 auto-switching AC. HART digital communications is a standard feature.
The unit supports SMART pH sensors by Rosemount Analytical. All configurations meet UL approval requirements for Ordinary Locations.
Special Features
- Full color display: High resolution full-color display (480 x 272 pixels)
- Local Operator Interface: Computer-like screens allow easy navigation for local configuration and routine calibration.
- Diagnostics: Continuous monitoring of the analyzer and the sensor loop(s) for fault and warning conditions.
- Data Logger: Records data from both channels and temperature every 30 seconds for 30 days
- Event Logger Audit Trail: Up to 300 significant events are recorded
- PID Control: Proportional, Integral and Derivative settings allow up to four analog current outputs to adjust a control device that has continuous adjustability
- Real-time Clock: Applies time/data stamps to all process data and events
- USB Data Port: Allows local transfer of all process data and events for evaluation on a computer
- Timer Functions: Time-proportional control, interval timers and four special application timers: bleed and feed, day and time interval timers, delay timer and a flow totalizer.
- Security Access: Two levels of password access
- Diagnostics: Continuously monitors and records the analyze and sensor(s) for fault conditions
- Wireless-ready: HART versions enable wireless transmissions of process variables and diagnostics from hard-to-reach locations with the Wireless THUM Adaptor.
- Case: Polycarbonate
- Dimensions: 6.2 x 6.2 x 5.2 in. (157 x 157 x 132mm)
- Conduit openings: Accepts (6) PG13.5 or 1/2 in. conduit fittings
- Display: Large 3.75 x 2.2 in. (95.3 x 55.9mm) high resolution
color LCD displays large process variables and user-definable display of diagnostic parameters. Calibration, programming and information screens display clear, easy-to-read characters. The color display is back-lit and backlighting intensity is user adjustable. Measurement character height: (.5") 13mm. Main display can be customized to meet user requirements.
- Ambient temperature and humidity: -10 to 60°C, (14 to 140°F) RH 5 to 95% (non-condensing). For Turbidity only: 0 to 55°C (32 to 131°F). RH 5 to 95% (non-condensing).
- Note: Some degradation in display response or performance
may occur below -5°C (23°F) and above 55°C (131°F). Above 60°C, the following components will progressively and automatically shut down: display, USB communications port, current outputs, alarm relays, main circuit board.
- WARNING: Always remove USB memory device if ambient temp exceeds 60°C. Do not access USB port if combustible atmosphere is present.
- Storage temperature: -20 to 60°C, (-4 to 140°F)
- Power:
- Code -02: 20 to 30 VDC. 20 W
- Code –03: 85 to 264 VAC, 47.5 to 65.0 Hz, 20 W
- Real time clock back-up: 24 hours.
- RFI/EMI: – EN-61326
- LVD: – EN-61010-1
- Model 56 Hazardous Location Approvals
- Options for CSA: -02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32,
34, 35, 36, 37, 38, HT and DP.
- Class I, Division 2, Groups A, B, C, & D
- Class Il, Division 2, Groups E, F, & G
- Class Ill T4A Tamb= 50°C
- Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indicators
adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S. respectively
- Options for FM: -02, 03, 20, 21, 22, 23, 24, 25, 26, 27, 30, 31, 32, 33, 34, 35, 36, 37, 38, HT and DP.
- Class I, Division 2, Groups A, B, C, & D
- Class Il & lll, Division 2, Groups E, F, & G
- T4 Tamp = -10 deg C to 60 deg C.
- Ordinary Locations (only with -UL ordering option):
- Options for UL: -02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32,
34, 35, 36, 37, 38, HT and DP.
- Altitude: for use up to 2000 meter (6562 ft.)
- Pollution Degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
- Altitude: for use up to 2000 meter (6562 ft.)
- Input: One or two isolated sensor inputs. Measurement choices
of pH/ORP, resistivity/conductivity/TDS, % concentration, ratio conductivity, total and free chlorine, monochloramine, dissolved oxygen, dissolved ozone, turbidity, pulse flow, temperature and raw 4-20mA input. For contacting conductivity measurements, temperature element must be a Pt1000 RTD. For other measurements (except ORP, flow and turbidity), use either a PT100 RTD, PT1000 RTD, or 22k NTC (D.O. only).
- Outputs: Four actively powered 4-20 mA or 0-20 mA isolated current outputs. Fully scalable. Max Load: 550 Ohms. Outputs can be
programmed for PID control. Output dampening can be enabled with time constants from 0 to 999 seconds. HART digital communications which is superimposed on output 1 is standard on all units (option code –HT).
- Alarms: Four alarm relays for process measurement(s) or
temperature. Any relay can be programmed for any measurement, timer, TPC or fault alarm operation, instead of a process alarm. When selected, a fault alarm will activate the relay when a sensor or analyzer fault occurs. Each relay can be configured independently. Alarm logic (high or low activation or USP) and deadbands are user-programmable.
- Relays: Form C, SPDT, epoxy sealed
- Inductive load: 1/8 HP motor (max.), 115/240 VAC
- Terminal Connections Rating
- Power connector (-02 24VDC power supply and -03 85-264VAC
power supply): 24-12 AWG wire size.
- Signal board terminal blocks: 26-16 AWG wire size.
- Current output connectors: 26-16 AWG wire size.
- Alarm relay terminal blocks: 24-12 AWG wire size.
- Weight/Shipping Weight: (rounded up to nearest lb or nearest
0.5 kg): 3 lbs/4 lbs (1.5 kg/2.0 kg) |
|
|