GENERAL
Sample requirements:
Pressure: 3 to 65 psig (122 to 549 kPa abs)
Inlet check valve opens at 3 psig (122 kPa abs).
Without check valve pressure is 1 psig (108 kpa abs)
Temperature: 32 to 122°F (0 to 50°)
Flow: 2-80 gal/hr (7.6-303 L/hr)
Sample Conductivity: >10 μS/cm
Process connection: 1/4" OD tubing compression fitting
Drain connection: 3/4" barbed fitting. Sample must drain to open atmosphere.
Wetted parts: acrylic, polycarbonate, polyester, Kynar1, nylon, silicone, PVC,
Viton2, silicone, polyethersulfone, polyester, copper, Tefzel2, glass, ceramic
Response to step change in chlorine: <120 sec to 90% of final reading
for inlet sample flow of 2 gph (7.6 L/hr)
Weight/shipping weight:
Model FCLi-01: 10 lb/13 lb (4.5 kg/6.0 kg)
Model FCLi-02: 11 lb/14 lb (5.0 kg/6.5 kg)
SENSOR
Free chlorine range: 0 to 20 ppm as Cl2. For higher ranges, consult the factory.
Accuracy: Accuracy depends on the accuracy of the chemical test used to calibrate the sensor
Linearity (0-20 ppm): 1% per IEC 60746
Linearity (0-2 ppm): ±0.05 ppm following calibration at 2 ppm
Sensitivity to pH: Between pH 6.5 and 10, sensor signal changes <4% per unit change in pH. Below pH 6.5 the change is <1%. per unit change in pH.
Interferences: Monochloramine, dichloramine, and permaganate
Electrolyte life: 3 months (approx.)
1056 ANALYZER
Case: Polycarbonate NEMA 4X/CSA 4 (IP65).
Display: Monochromatic back-lit LCD. Main character height 0.6 in (15mm). Display is user-programable
Languages: English, German, Italian, Spanish, French, Portuguese
Ambient temperature and humidity: 32 to 131°F (0 to 55°C); RH 5 to 95% (con-condensing)
Storage temperature: -4 to 140°F (-20°C and 60°C)
Power: 85 to 265 Vac, 47.5-65.0 Hz 15 W. Equipment protected throughout by double insulation.
RFI/EMI: EN-61326
LVD: EN-61010-1
Outputs: Two 4-20 mA or 0-20 mA isolated outputs. Continuously adjustable. Linear or logarithmic. Maximum load 550 ohms. Output dampening with time constant of 5 sec is userselectable.
Alarms: Four alarm relays for chlorine, pH or temperature. Fully programmable. Any alarm can be configured as a fault alarm.
Relays: Form C, SPDT, epoxy sealed
54EA ANALYZER
Case: Epoxy-painted cast aluminum, NEMA4X (IP65).
Display: Three-line, back-lit, dot matrix LCD. First line is measurement reading. Second line is temperature and current output. Third line is user-selectable.
Character heights: 1st line - 0.6 in. (16 mm) , 2nd and 3rd lines - 0.3 in. (7 mm) .
Ambient Temperature and Humidity: 0 to 50°C (32 to 122°F). 95% (maximum) non-condensing. Analyzer can be operated between -20 and 60°C (-4 to 140°F) with some degradation in display quality.
Power: 100-127 Vac ± 10%, 50/60 Hz ± 6%, 8 W
200-253 Vac ± 10%, 50/60 Hz ± 6%, 8 W
RFI/EMI: EN-61326
LVD: EN-61010-1
Outputs: Two 4-20 mA or 0-20 mA isolated outputs. Continuously adjustable. Output dampening is user-selectable. Maximum load at 115/230 Vac is 600Ω. Maximum load at 100/200 Vac is 550Ω.
Alarms: Three alarm relays for chlorine, pH or temperature. Fully programmable. Fourth relay for analyzer/sensor fault.
Relays: Relays 1-3: Form A, SPST, epoxy sealed
Relay 4: Form C, SPDT, epoxy sealed
Resistive Inductive
28 Vdc 5.0 3.0
115 Vac 5.0 3.0
230 Vac 5.0 1.5