SPECIFICATIONS - GENERAL
Case: Polycarbonate (panel and surface/pipe mount). NEMA 4X/CSA 4 (IP65)
Dimensions Panel (code -10): 6.10 x 6.10 x 3.72 in. (155 x 155 x 94.5 mm)
Surface/Pipe (code -11): 6.23 x 6.23 x 3.23 in. (158 x 158 x 82 mm)
Conduit openings: Accepts PG13.5 or 1/2 in. conduit fittings
Ambient Temperature: 32 to 122°F (0 to 50°C). Some degradation of display above 50°C.
Storage Temperature: -4 to 158°F (-20 to 70°C)
Relative Humidity: 10 to 90% (non-condensing)
Weight/Shipping Weight: 2 lb/3 lb (1 kg/1.5 kg)
Display: Two line, 16-character display. Character height: 4.8 mm; first line shows process variable (pH, ORP, conductivity, % concentration, oxygen, ozone, chlorine, or monochloramine), second line shows process temperature and output current. For pH/chlorine combination, pH may also be displayed. Fault and warning messages, when triggered, alternate with temperature and output readings. During calibration and programming, messages, prompts, and editable values appear on the two-line display.
Temperature resolution: 0.1°C (≤99.9°C); 1°C (≥100°C)
Hazardous Location Approval: For details, see specifications for the measurement of interest.
RFI/EMI: EN-61326
DIGITAL COMMUNICATIONS:
HART —
Power & Load Requirements: Supply voltage at the transmitter terminals should be at least 12 Vdc. Power supply voltage should cover the voltage drop on the cable plus the external load resistor required for HART communications (250 Ω minimum). Minimum power supply voltage is 12 Vdc. Maximum power supply voltage is 42.4 Vdc.
Analog Output: Two-wire, 4-20 mA output with superimposed HART®3 digital signal. Fully scalable over the operating range of the sensor.
Output accuracy: ±0.05 mA
FOUNDATION FIELDBUS —
Power & Load Requirements: A power supply voltage of 9-32 Vdc at 13 mA is required. Fieldbus Intrinsically Safe COncept/FISCOcompliant versions of Model Xmt Foundation Fieldbus transmitters are available.