Power:
• Standard: 24 VDC (20-30 VDC)
• Optional: 90-264 VAC, 47-63 Hz
Environmental temperature:
-20° to 60° C (-4° to 140° F)
Environmental temperature without safety certification:
-40° to 60° C (-40° to 140° F)
Enclosure Protection Rating: IP66 and NEMA 4X
Dimensions (without sample system):
• Wall-mount:
711.20 mm H x 444.5 mm W x 497.84 mm D
(28” H x 17.5” W x 19.6” D)
• Pipe-mount:
711.20 mm H x 444.5 mm W x 670.56 mm D
(28” H x 17.5” W x 26.4” D)
• Floor-mount:
1531.62 mm H x 444.5 mm W x 612.14 mm D (60.3” H x 17.5” W x 24.1” D)
Mounting: Free-standing (standard), wall- or pipe-mount (optional)
Approximate Weight (without sample system): 49.895 kg (110 lbs.)
Area Safety Certification Options:*
• CSA:
– USA
• Class I, Zone 1, A Ex d IIC, T6
• Class I, Division 1, Groups B, C, D, T6, Enclosure Type 4
– Canada
• Class I, Zone 1, Ex d IIC, T6
• Class I, Zone 1, Ex d IIC, T6
• ATEX/IECEX
– Ex II 2G
– Ex d IIC Gb T6
Oven: Airless heat sink, maximum 150° C (302° F)
Valves: 6-port and 10-port diaphragm chromatograph valves. Other types of valves, such as liquid injection or rotary valves, may be used depending on the application
Carrier Gas: Application-dependent. Typically zero-grade helium, nitrogen, or hydrogen
Sample Input Pressure Range (recommended): 15-20 psig
Carrier Gas Input Pressure Range (recommended): 90-100 psig
Detector: Thermal conductivity detector (TCD), flame ionization detector (FID), TCD/TCD or TCD/FID dual detector configurations possible; flame photometric detector (FPD) available (see FPD module data sheet)
Gating Options: Fixed-time, slope sensing gating of peaks
Streams: Up to 20 streams (including calibration stream), 6 streams standard
Chromatograms stored/archived internally: Stores up to 30 days of analysis report data and up to 2500 individual chromatograms.
Communications (Standard):
Ethernet: Two available connections – one RJ-45 port & one 4-wire termination – with 10/100mbps
Analog inputs: Two standard inputs filtered with transient protection, 4-20mA (user scalable and assignable)
Analog outputs: Six isolated outputs, 4-20mA
Digital inputs: Five inputs, user assignable, optically isolated, rated to 30VDC @ 0.5A
Digital outputs: Five user-assignable outputs, Form C and electromechanically isolated, 24VDC
Serial: Three termination blocks, configurable as RS-232, RS-422 or RS-485 and one D-sub (9-pin) port for PC connection
Communications (Options):
Two expansion slots available for additional communications. Each slot has the capacity to add one of the following:
4 analog inputs (isolated) card
4 analog outputs (isolated) card
8 digital inputs (isolated) card
5 digital outputs (isolated) card
1 RS-232, RS-422 or RS-485 serial connection card
1 modem card, 300-19.2k baud
Memory Capacity: 1 Gb of flash memory for data storage; 128Mb of SDRAM system memory with 2 Mb static RAM (battery-backed)
*Use of the optional LSIV will result in slightly different certifications.