7827 High Performance Viscosity Meters
The 7827 digital viscosity meters is used for quality and analytical measurements where kinematic viscosity at line and reference temperature are required to be measured. The essential accuracy and speed of response makes this the first true viscometer solution for process measurement. When 7827 digital viscosity meters are used in conjunction with 795x Series Signal Converters, a variety of referral methods are avialable to compute the viscosity at the reference temperature from the viscosity measured at the process conditions.
Transmitter Feature Comparison Chart
|
|
- Continuous multivariable measurement of viscosity, density & temperature
- 7827 long stem viscosity meter is available in lengths up to 13ft (4m) and is suitable for open and closed tanks
- Wetted parts 316L stainless steel, nickel alloy, Monel, Titanium
- Pressure rating to 3000 psi (207 bar) for short stem, 1450 psi (100 bar) for long stem versions
- Temperate range (-60°F to +392°F (-50°C to +200°C))
- Viscosity accuracy ±1% full scale
- Density accuracy ±0.001g/cc (±1.0kg/m3) (±0.06 lb/ft3)
- Hazardous area approved (CSA and ATEX)
- Ideal for Newtonian fluids
- Direct line viscosity, density & temperature measurement
- Rugged and reliable, direct insertion meters with very low maintenance to deliver confidence and savings
- Viscosity and temperature signals enable on-line measurements of referred viscosity, viscosity index and ignition index (CII / CCAI)
- Density and temperature signals enable a range of parameters to be measured including specific gravity, % solids, % volume concentration and oBrix delivering flexibility for your operation
- Range of corrosion-resistant materials, process connections, and stem lengths deliver just what is required for the application
- Optimized design that is insensitive to vibration, temperature and pressure variations
|
Brief Specifications* |
|
Viscosity accuracy |
±1% full scale calibrated range |
|
Viscosity range |
0-20,000cP |
|
Repeatability |
±0.5% reading |
|
Density accuracy |
±0.001g/cc (±1.0kg/m3) (±0.06 lb/ft3) |
|
Density range |
0 - 3g/cc (0 - 3000kg/m3) (0-187.4 lb/ft3) |
|
Repeatability |
±0.0001g/cc (±0.1kg/m3) (±0.006 lb/ft3) |
|
Temperature effect |
±0.0001 g/cc/°C |
|
Pressure rating |
Short stem: 207bar (3000psi) Long stem: 100bar (1450psi) |
|
Temperature range |
Process -60°F to +392°F (-50°C to +200°C)
Ambient -40°F to +185°F (-40°C to +85°C) |
|
Wetted parts |
Stainless Steel 316L, Nickel Alloy, Monel 400, Titanium, Zirconium |
|
Stem Length Range |
1.6 – 13ft (0.5 – 4m) |
|
Electronics |
Frequency output |
|
* Listed specifications may vary with model. |
- Turbine meter viscosity correction
- Fuel oil blending
- Power station burner control
- Lubricating oils
- HFO bunker blending
|
|
|