Power: 115 VAC ±15%, 220 VAC 15%, 50/60 HZ100 watts running, 350 watts start-up
Environment: -18° to 55°C (0° to 130°F)
Enclosure Protection Rating: NEMA 4
Dimensions (without sample system):
164cm H x 51cm W x 53.5cm D
(64.5” H x 20” W x 21” D)
Mounting: Free-standing (standard)
wall-mount (optional)
Approximate Weight (without sample system): approx. 36 kg. (79 lbs.)
Area Safety Certification Options (hardware dependant)
Standard: Designed to meet UL Class 1, Div. 1, Groups C, D (Group B with optional air purge)
Optional CSA NRTL/C Certified: Class I, Div. 1, Groups C, D, T3B
Optional ATEX Certified: EEx d IIB T4amb = 60°C CE marked per ATEX Directive (94/9/EC)
Oven: Airless heat sink, maximum 85°C (185°F)
Valves: 6-port and 10-port diaphragm chromatograph valves. Other types of valves may be used depending on the application such as liquid injection and rotary valves
Carrier Gas: Application dependent. Typically zero-grade helium, nitrogen or hydrogen at 90 psig
Detector: Thermal Conductivity Detector (TCD); up to two depending on application; Flame Photometric Detector (FPD) available (see FPD Module Data Sheet)
Gating Options: Fixed-time, slope and automatic gating of peaks
Streams: Up to 12 streams
Chromatograph Control Electronics: Mounted with the gas chromatograph in explosion-proof housing or remotely in 19” rack-mounting
Analog Inputs: Four inputs filtered with transient protection (note that the 4 inputs will be used by the second TCD)
Analog Outputs: Two outputs standard (up to 10 optional), 4 - 20 mA, non-isolated
Serial Communication Ports: Three serial ports standard with option for a total of eight. Depending on the port, choice of RS- 232, RS-422 and RS-485 is available as well as the Modbus protocol
Digital Inputs: One gas chromatograph alarm and four user assignable inputs, optically isolated with transient protection
Digital Outputs: Five digital outputs can be used for alarms, optically isolated with transient protection
Parallel Printer Port: One parallel port available for printed reports
Internal Modem (optional): Field-configurable; 300 to 19.2k baud
Transient Protection: C.E. tested and certified to the highest levels (3 and 4) of the European IEC 801 STD