Power:
Standard: 24 VDC, (24.5-26.5 VDC)
Optional: 90-264 VAC, 47-63 Hz
80 watts start-up; 33 watts nominal
Environment: -29° to 60°C (-20° to 140°F)
Enclosure Protection Rating: IP66 and NEMA 4X
Dimensions (without sample system):
Wall-mount: 914mm H x 462mm W x 483mm D
(36” H x 18.2” W x 19” D)
Pipe-mount: 914mm H x 462mm W x 645mm D
(36” H x 18.2” W x 25.4” D)
Floor-mount: 1483mm H x 419mm W x 577mm D
(58.4” H x 16.5” W x 22.7” D)
Mounting: Free-standing (standard); wall-mount or pipe-mount (optional)
Approximate Weight (without sample system): 80 kg (175 lbs.)
Area Safety Certification Options:
CSA: Class 1, Division 1, Groups B, C, D, T4
USA: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4
Canada: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4
ATEX: Ex II 2G Ex d IIC T4 Tamb = 60°C
Oven: Airless heat sink, maximum 130°C (266°F)
Valves: 6-port and 10-port diaphragm chromatograph valves. Other types of valves may be used, such as liquid injection or rotary valves, depending on the application
Carrier Gas: Application-dependent. Typically zero-grade helium, nitrogen, or hydrogen
Sample Input Pressure Range (recommended): 15-20 psig
Carrier Gas Input Pressure Range (recommended): 90-100 psig
Detector: Thermal conductivity detector (TCD), flame ionization detector (FID), TCD/TCD or TCD/FID dual detector configurations possible; flame photometric detector (FPD) available (see FPD module data sheet)
Gating Options: Fixed-time automatic slope sense gating of peaks
Streams: Up to eight streams (including calibration stream) internally controlled; up to 12 streams externally controlled
Chromatograms stored/archived internally: Most recent analysis and/or calibration per stream
Analog Inputs: Four inputs filtered with transient protection (user scalable and assignable)
Analog Outputs: Four non-isolated analog outputs, 4 – 20 mA (standard); four additional isolated analog outputs or eight additional non-isolated analog outputs (optional)
Serial Communication Ports: Three serial ports standard with option for a total of seven. Depending on the port, choice of RS-232, RS-422, and RS-485 is available as well as the Modbus protocol
Digital Inputs: Five digital inputs; user assignable
Digital Outputs: One gas chromatograph alarm and four user-assignable outputs (2 form A, 3 solid-state optically isolated)
Parallel Printer Port: One parallel port available for printed reports
Internal Modem (optional): Field-configurable; 300 to 33.6k baud rate
Ethernet Card (optional): 10 mbps with RJ-45 port
RAM (DATA) Memory Capacity: 4 Mb RAM (standard); additional 16 Mb Flash memory (optional)