- Electrical Power: 115 VAC +/- 15% at 0.55 A, 50/60 Hz;
- 230 VAC +/- 15% at 0.275 A, 50/60 Hz
-
- Communications:
- Four serial ports standard (user-selectable RS-232/484/422)
- Optional:
- - Com 4A board for additional serial ports (5, 6, 7, 8)
- - Internal modem connected to dedicated serial port
- - Ethernet network interface card
-
- Analog Inputs:
- • Detector 1-4 each filtered with transient protection dedicated to the
- GC auto-ranging pre-amp.
- • Detector 2-4 each filtered with transient protection for optional
- inputs from other devices or for use with a second auto-ranging pre-
- amp from a second GC detector. 16 bit A/D resolution.
-
- Analog Outputs:
- Two outputs standard
- 4-20 mA with software calibration
- Optional: 4 or 8 additional outputs
-
- Digital Inputs:
- • One GC alarm, optically isolated with transient protection
- • Five user alarms, optically isolated, with transient protection
- • 12 V standard, maximum 24 V external
-
- Digital Outputs: Five digital outputs can be used for alarms, optically isolated, with transient protection. 30 VAC 500 mA maximum
Keypad Display: Optional 18 keys/8 lines x 41 characters. Uses Com 4 or Com 8, depending on serial port options.
Parallel Port: One parallel port available for printed reports.
Internal Modem: Optional 300 to 33.6 k baud field configurable.
Detector Inputs: TCD, FID, FPD, single or dual pre-amp inputs.
Transient Protection: CE tested and certified to the highest levels (3 & 4) of the European IEC 801 STD.
Physical Environment: Operating range at 0° to 130° F
(-18° to 55° C), storage range at -40° to 185° F (-40° to 85° C), Humidity 0-95% RH non-condensing.
- Mounting Options:
- • Integral/field mount - X-proof - 60 lbs (27.21 kg) NEMA 4X, IP 65 19
- • Rack-mount - 25 lbs (11.34 Kg)
- • Panel-mouth retrofit - 22 lbs (9.98 Kg)
System Memory: 16 MB disk on a chip storage for all historical data (expandable up to 32 MB historical storage space).
Chromatogram Storage: Last run for each stream and last call for each method. Additional chromatogram storage in MON2000 software running on a computer.
Methods: Four timed event tables, four component tables freely assignable to each stream.
- Safety Classification:
- 2350A Explosion Proof Controller
- UL-listed for use in Class l, Division 1, Groups C and D hazardous locations both the US and Canada. ATEX-certified to EEx d llB T6 standards EN50014 and EN50018 for potentially explosive atmoshperes Parts 1 and 5.
-
- 2350A Rack Mount
- CSA-certified