- Power: 115 VAC ±15%, 220 VAC ±15%, 50/60 HZ
- 400 watts running, 1100 watts start-up
-
- Environment: 0° to 55°C (32° to 130°F) for TCD and FID;
- 20°C ± 5°C (70°F ± 10°F) for FPD
-
- Dimensions (without sample system):
- 198cm H x 61cm W x 61cm D (78" H x 24" W x 24" D)
Mounting: Wall-mounted (standard); free-standing (optional)
- Approximate Weight (without sample system):
- Approximately 68 kg. (150 lbs.)
Area Classification Options (Hardware dependent): designed to meet Class 1, Div. I, Groups B, C, D with X-purge; Class 1, Div. 2, Groups B, C, D with Z-purge, ATEX certified EExpII T3 (optional)
Instrument Air: 4 SCFM at 40 psig of oil-free instrument air for purge and oven heat and 90 psig for valve actuation
Oven: Air bath oven, maximum 150°C (300°F)
Valves: 6-port and 10-port diaphragm chromatograph valves (maximum 6 valves). Other types of valves may be used depending on the application such as liquid injection and rotary valves
Carrier Gas: Application-dependent. Typically zero-grade helium, nitrogen or hydrogen at 90 psig
Columns: choice of packed, micro-packed or capillary columns; application-dependent
Detector: Thermal conductivity detector (TCD), flame ionization detector (FID) or flame photometric detector (FPD); TCD-TCD, TCD-FID or TCD-FPD dual detector combinations possible
Gating Options: Fixed-time automatic slope sense gating of peaks
Streams: up to 12 streams
Chromatograph Control Electronics: Integrally mounted
Analog Inputs: Four inputs filtered with transient protection (user scalable and assignable)
Analog Outputs: Four non-isolated analog outputs, 4 - 20 mA (standard); four additional isolated analog outputs or eight additional non-isolated analog outputs (optional)
Serial Communication Ports: Three serial ports standard with option for a total of seven. Depending on the port, choice of RS-232, RS-422, and RS-485 is available as well as the Modbus protocol
Digital Inputs: Five digital inputs; user assignable
Digital Outputs: One gas chromatograph alarm and four user-assignable outputs (2 form A, 3 solid-state optically isolated)
Parallel Printer Port (optional): one parallel port available for printed reports
Internal Modem (optional): Field-configurable; 300 to 19.2k baud