- Power: 115 VAC ±15%, 220 VAC ±15%, 50/60 HZ
- 100 watts running, 350 watts start-up
Environment: -18° to 55°C (0° to 130°F)
Enclosure Protection Rating: NEMA 4
- Dimensions (without sample system):
- 164cm H x 51cm W x 53.5cm D (64.5" H x 20" W x 21" D)
Mounting: Free-standing (standard); wall-mount (optional)
- Approximate Weight (without sample system):
- Approximately 36 kg. (79 lbs.)
Area Safety Certification Options (hardware dependent):
Standard: Designed to meet UL Class l, Div. 1, Groups C, D (Group B with optional air purge)
Optional CSA NRTL/C Certified: Class l, Div. 1, Groups C, D T3B
- Optional ATEX Certified: EEx d llB T4amb = 60°C
- CE marked per ATEX Directive (94/9/EC)
Oven: Airless heat sink, maximum 85°C (185°F)
Valves: 6-port and 10-port diaphragm chromatograph valves. Other types of valves may be used depending on the application such as liquid injection and rotary valves
Carrier Gas: Application dependent. Typically zero-grade helium, nitrogen, or hydrogen at 90 psig
Detector: Thermal conductivity detector (TCD); up to two depending on application. Flame photometric detector (FPD) available (see FPD Module data sheet)
Gating Options: Fixed-Time, Slope, and Automatic gating of peaks
Streams: Up to 12 streams (including calibration stream)
Chromatograph Control Electronics: Mounted with the gas chromatograph in explosion-proof housing or remotely in 19" rack-mounting
Analog Inputs: Four inputs filtered with transient protection (note that the 4 inputs will be used by the second TCD)
- Analog Outputs: Two outputs standard (up to 10 optional), 4-20 mA,
- non-isolated
Serial Communication Ports: Three serial ports standard with option for a total of eight. Depending on the port, choice of RS-232, RS-422, and RS-485 is available as well as the Modbus protocol
Digital Inputs: One gas chromatograph alarm and four user assignable inputs, optically isolated with transient protection
Digital Outputs: Five digital outputs can be used for alarms, optically isolated with transient protection
Parallel Printer Port (optional): One parallel port available for printed reports
- Internal Modem (optional): Field-configurable; 300 to 19.2k baud
-
- Transient Protection: C.E. tested and certified to the highest levels (3 and 4) of the European IEC 801 STD